Care instructions ESTETICA E30
3 Replenishing and checking the disinfectant | 3.1 Replenishing KaVo OXYGENAL 6 when using the water bottle
▶
Squeeze the KaVo OXYGENAL 6 bottle until the germ reduction liquid rises
above the red marker.
ð
In response to over-filling, the excess quantity above the tick mark is aspir-
ated automatically to ensure exact dosing.
3.1.1 Filling the intensive germ reduction bottle
CAUTION
Wrong germ reduction liquid or the dose is too high
Injury hazard
▶
Use KaVo OXYGENAL 6 only.
▶
Do not exceed the maximum concentration of 0.25% KaVo OXYGENAL 6.
▶
Proceed like for filling the water bottle, but use a different dose of KaVo
OXYGENAL 6.
See also:
2
Filling the water bottle with the dosing facility, Page 36
▶
Fill the four-fold dose of KaVo OXYGENAL 6 (equivalent to 20 ml) in the in-
tensive germ reduction bottle. Make sure that the spout of the KaVo OXY-
GENAL 6 dosing facility (nose) is situated right above the opening of the in-
tensive germ reduction bottle.
▶
Then fill-up the intensive germ reduction bottle with tap water.