7.4
Combined heat/cooling metering
MULTICAL
®
402 is available as heat meter (Meter type 2xx), cooling meter (Meter type 5xx) or combined
heat/cooling meter (Meter type 6xx).
Meter type
Heat meter (MID)
2
Cooling meter
5
Heat/cooling meter
6
Country code (language on label etc.)
XX
If MULTICAL
®
402 has been supplied as a combined heat/cooling meter (meter type 6xx), it measures heat energy
(E1) at a positive temperature difference (T1
>
T2), whereas it measures cooling energy (E3) at a negative
temperature difference (T2
>
T1). Temperature sensor T1 (with a red type sign) must be installed in the hydraulic
forward pipe, whereas T2 (with a blue type sign) is installed in the return pipe.
If the current T1 exceeds, or equals
θ
hc
only heat energy can be measured. If the current T1 is lower than or equals
θ
hc
θ
only cooling energy can be measured.
hc
is the temperature point used to change between heat and cooling measurement.
θ
hc
In combined heat/cooling meters
θ
is configurable in
temperature range 0.01…160.00
°
C.
hc
should correspond to the highest occurring forward temperature in
connection with cooling, e.g. 25
°
C. If the meter is to be used for ”purchase and sale of heat”,
θ
hc
is set at
180.00
°
C, which cancels the
θ
hc
If you want to switch the qhc function on or off compared to current condition, it is necessary to perform a total
programming of the meter by means of METERTOOL.
function.
The change between heat and cooling measurement involves no hysteresis (
∆θ
hc
θ
= 0.00K).
hc
is configured by means of METERTOOL (see paragraph 14.2).
Summary of Contents for MULTICAL 402
Page 1: ...MULTICAL 402 Technical Description...
Page 2: ...MULTICAL 402 2 Kamstrup A S Technical Description 5512 742_M1_GB_01 2016...
Page 109: ...MULTICAL 402 Kamstrup A S Technical Description 5512 742_M1_GB_01 2016 109...
Page 113: ...MULTICAL 402 Kamstrup A S Technical Description 5512 742_M1_GB_01 2016 113...