MULTICAL® 402
Kamstrup A/S · Technical Description · 5512-742_M1_GB_01.2016
105
Example: A MULTICAL
®
1% of qp:
402 flow sensor shows the following result after verification:
+1.1%
10% of qp:
+0.3%
100% of qp:
-0.1%
In order to correct the inaccuracies, the following values are entered:
1% of qp:
-1.1%
10% of qp:
-0.3%
100% of qp:
+0.1%
Adjustments of more than +/-5% ought not to be made, as they can be due to a flow sensor error.
Having been adjusted the flow sensor is now ready for test/verification as well as sealing.
Note!
Current handling and reverification requirements must be observed.