HygroPort I.S. User’s Manual
Appendix C
Datalog Status Display
Introduction
When the status of the current log file is displayed, see
Figure C.1, the
ST
column in
the current log file display represents the instrument’s status at the time that each log
point is recorded.
File
CR270309
INT
RH
EXT.
EXT_DP
INTVAL
05 Sec
D#
090327
T#
123606
Figure C.1
Current datalog file display
This column shows a hexadecimal representation of the instrument’s 16-bit status
register shown below in Figure C.2. Bit 1 is the least significant bit. The four most
significant bits, Bits 13 to 16 are for Kahn Instruments’ service use and so the first
character in any HygroPort I.S. ST display will always read 0.
Status register Bit position
12
11
10
9
8
7
6
5
4
3
2
1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
N/A
Figure C.2
HygroPort I.S. status register
For the example given for each of the log plots shown on in Figure C.1 (Hex 0028), the
output pattern of the status register will be as shown in Figure C.3
.
Status register Bit position
12
11
10
9
8
7
6
5
4
3
2
1
0
0
0
0
0
0
1
0
1
0
0
N/A
Figure C.3
HygroPort I.S. status register (Hex 28)
Each of the error bits set high (1) represents a specific status or error condition. In
this example, Bits 4 and 6 are set high indicating that two (status) conditions exist.
Issue 6, December 2019
55
Kahn Instruments
INT
26.8
26.2
25.9
25.8
25.5
ST
0028
0028
0028
0028
0028
EXT
-
100.0
-90.0
-80.0
-70.0
-60.0
#
106
107
108
109
110
16
15
14
13
N/A
N/A
N/A
N/A
16
15
14
13
N/A
N/A
N/A
N/A