HygroPort I.S. User’s Manual
7.3
Calibration Correction Method
To enter the Calibration Menu screen, highlight
CALIBRATION
from the Main Menu by
using a combination of the scroll up and scroll down keys and press the Enter key.
NOTE: Any logging procedure currently in operation must be cancelled
before the calibration procedures can be entered.
Entry to the Calibration Menu is protected by a PIN which must first be entered. The PIN
(
4876
) is entered in a similar manner to that described in Section 2.5.
This gives access to the HygroPort I.S. calibration look-up table as shown in Figure 7.2
and provides the facility to select and change any single or number of points by means
of the scroll up, scroll down, scroll right and scroll left keys.
POINT
REF
MDM300
-100
-100.0
-100.0
-90
-90.0
-90.0
-80
-80.0
-80.0
-70
-70.0
-70.0
-60
-60.0
-60.0
Figure 7.2
HygroPort I.S. calibration correction screen
Issue 6, December 2019
45
Kahn Instruments
CALIBRATION