HygroPort I.S. User’s Manual
2.6.2 Logging
Highlight
LOGGING
from the Main Menu by
using a combination of the
(up) and
(down) keys and press the Enter key. The
Logging Set-up menu, is then displayed, see
Figure 2.9.
Initially the
FILE NAME
fi ield is highlighted. This
fi ield is used to defi ne a name for a new log file
and can be any combination of Alphanumeric
characters (max 8 digits). NOTE: If a file
name is not entered, a logging cycle
cannot be started.
To enter the parameter value fi ield press the
Enter key. A prompt (-) for the first character
is then highlighted. Use a combination of the
scroll up and scroll down keys to insert the first
character and then the scroll right key to shift to
the second character position.
Repeat the above procedure until all characters
have been entered and press the Enter key.
Set up the
PARAMETER
to be logged and the
LOG INTERVAL
in a similar manner.
After setting up the logging parameters, start the log cycle by selecting
START?
. Press
the Enter key to highlight
STOPPED
and use the scroll down key to change the
field value to
STARTED
. Press the Enter key to start the logging cycle.
Table 2.4 details the field values. To quit, press the scroll left key.
Parameter variable options
Defines datalog file name. Field width up to eight alpha- numeric
characters. Set with the scroll up, scroll down, scroll right and
scroll left keys (NOTE: Scroll left
deletes characters)
Sets parameter to be logged. This need not necessarily be the
same as that displayed on the Main Display. Options available,
DEWP, GM
3
, GKG, RH, PPMW, PPMV, CALC DP
Sets Log Interval in 5 sec increments. Range 5 to 600 seconds
Controls datalog Start and Stop functions. Logging cannot be
started if a file name has not been defined. Once started, log file
parameter fi elds are locked out and cannot be changed.
Table 2.4
Logging parameter set up
Issue 6, December 2019
15
Kahn Instruments
FILE NAME
-
>
LOGGING
PARAMETER
DEWP
LOG INTERVAL
5
s
START?
STOPPED
FILE NAME
-
>
LOGGING
PARAMETER
DEWP
LOG INTERVAL
5
s
START?
STOPPED
FILE NAME
N2A-
>
LOGGING
PARAMETER
DEWP
LOG INTERVAL
5
s
START?
STOPPED
FILE NAME
N2A
>
LOGGING
PARAMETER
DEWP
LOG INTERVAL
5
Sec
START?
STOPPED
Figure 2.9
Set-up Logging menu
Parameter
FILE NAME
PARAMETER
LOG
INTERVAL
START ?