40601200T10Z001K000
The jumpers for configuration are located on the backplane in the transmitter housing.
Configure the digital outputs as described below:
1.
Switch off the voltage supply and wait at least 20 minutes before performing the next step.
2.
Screw in the lid lock (4) and open the housing lid (1).
3.
Loosen the screws (3) and pull out the transmitter plug-in module (2).
4.
Insert the jumpers at the required positions.
5.
Re-insert the transmitter plug-in module (2) and retighten the screws (3).
6.
Close the housing lid (1) and secure the lid by unscrewing the screw (4).
HART protocol
PROFIBUS-PA
G
008
7
4-01
2
3
3
1
4
1
2
12
BR
9
02
BR
9
01
1
2
BR
9
02
BR902 for digital output DO1
BR901 for digital output DO2
BR902 in position 1: standard (non-NAMUR)
BR902 in position 1: standard (non-NAMUR)
BR902 in position 2: NAMUR
BR902 in position 2: NAMUR