Free inlet, free outlet
Heavily contaminated measurement media
Installation near pumps
•
For a free outlet, do not install the measuring device at the highest
point or in the draining side of the pipeline, as the measuring pipe
runs empty or air bubbles can form (1).
•
For free inlets or outlets, provide a culvert (downward routing of
the pipe) so that the pipeline is always full (2).
G
00040
(1)
(2)
•
We recommend a bypass connection, according to the figure, for
heavily contaminated media so that the system can continue op-
erating without any interruptions during mechanical cleaning.
G00042
•
We recommend using mechanical vibration compensators for
sensors that are installed near pumps or other equipment gener-
ating vibrations.