V1.00/EN/00643259
Page 34/57
Data Sheet 406012
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone:
+49 661 6003-0
Fax:
+49 661 6003-607
Email:
Internet:
www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:
+44 1279 635262
Email:
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866
1-800-554-5866
Fax:
315-437-5860
Email:
Internet: www.jumousa.com
40601200T10Z001K000
Model 406012/1-1, 406012/2-1 with 406018/2-1 in zone 1 - HART protocol and PROFIBUS-PA
Voltage supply
Signal cable connection
(for devices with remote mount design only)
Input and output connection
A = Transmitter, B = Sensor
SE
SE
31 32
A
24 V
5
1
5
2 81 82 41 42
L
N
1+ 2-
M1 M2 D1 D2
3
2S E2 E1 1S
M1 M2 D1 D2
3
E2 E1
B
PE
PA
+
-
+
-
+
-
+
-
HART
PA
97 9
8
41 42
PA+ PA-
+
-
PROFIBUS-PA
406012/1-1
406012/2-1
406018/2-1
G
0132
7
PA
PA
PA
Alternating voltage (AC)
Direct voltage (DC)
Terminal
Function
Terminal
Function
L
Phase
1+
+
N
Neutral conductor
2-
-
PE
Protection conductor (PE)
PE
Protection conductor (PE)
Terminal
Function
Wire color
M1
Magnet coil
Brown
M2
Magnet coil
Red
D1
Data line
Orange
D2
Data line
Yellow
SE
Shielding
-
E1
Signal line
Violet
1S
Shield of E1
-
E2
Signal line
Blue
2S
Shield of E2
-
3
Measurement potential
Green
Terminal
Function
31/32
Current/HART output
- The current output is available in "active" or "passive" mode. The configuration must be specified when or-
dering the unit, since the configuration cannot be changed on-site.
97/98
PROFIBUS-PA (PA+ PA-)
- According to IEC 61158-2
51/52
Digital output DO1 active/passive
- Function can be configured using software on-site as "Pulse output" or "Binary output". The
default setting is "Pulse output".
81/82
Digital input/contact input
- Function can be configured using software on-site as "External output switch-off", "External totalizer
reset", "External totalizer stop", or "Other". Only available in conjunction with "passive" current output.
41/42
Digital output DO2 passive
- Function can be configured using software on-site as "Pulse output" or "Binary output". The default
setting is "Binary output", flow direction signaling.
PA
Potential equalization (PA)