HT30-M1
63
Configuration
>Identif. off
Setpoint RS232
Autostart off
(**
An adjustment of these values via a serial interface only makes sense, if
the parameter for the identification is on >off<.
If the identification is adjusted (see page 36), the control parameters Xp,
Tn and Tv of a control course are determined and stored automatically.
Therefore the adjusted values are overwritten.
11.4. Status messages / error messages
The circulator sends data (including error messages) only when the computer sends a query.
Status messages
Description
00 MANUAL STOP
Circulator in "OFF" state.
01 MANUAL START
Circulator in keypad control mode.
02 REMOTE STOP
Circulator in "r OFF" state.
03 REMOTE START
Circulator in remote control mode.
Error messages
Description
-01 LOW LEVEL ALARM
Low liquid level alarm
-02 COOLING UNIT ALARM
Failure of one or both temperature sensors of the C.U.
cooling unit.
-03 EXCESS TEMPERATURE WARNING
High temperature warning "
".
-04 LOW TEMPERATURE WARNING
Low temperature warning "
".
-05 WORKING SENSOR ALARM
Working temperature sensor short-circuited or interrupted.
-06 SENSOR DIFFERENCE ALARM
Sensor difference alarm.
Working temperature and safety sensors report a
temperature difference of more than 100 K.
-07 I
2
C-BUS ERROR
Internal error when reading or writing the I
2
C bus.
-08 INVALID COMMAND
Invalid command.
-09 COMMAND NOT ALLOWED IN CURRENT
OPERATING MODE
Invalid command in current operating mode.
-10 VALUE TOO SMALL
Entered value too small.
-11 VALUE TOO LARGE
Entered value too large.
-12 TEMPERATURE MEASUREMENT ALARM
Error in A/D converter.
-13 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
Value lies outside the adjusted range for the high and low
temperature warning limits. But value is stored.
-14 TEMPERATURE/LEVEL ALARM
Safety temperature alarm