Remote control
62
out commands:
Setting parameters or temperature values.
Command
Parameter
Response of circulator
out_mode_02
0
No identification.
Temperature control by using the stored parameters.
out_mode_02
1
Single identification of controlled system after the next start.
out_mode_02
2
Continual identification of controlled system whenever a new setpoint is
to be reached.
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the circulator = r OFF.
out_mode_05
1
Start the circulator.
out_mode_08
0
Control dynamics aperiodic
out_mode_08
1
Control dynamics standard
out_sp_00
xxx.x
Set working temperature "T".
out_sp_03
xxx.x
Set high temperature warning limit
.
out_sp_04
xxx.x
Set low temperature warning limit
.
out_hil_00
xxx
Set the desired maximum cooling power (0 % to 100 %).
This adjustment is required only for proportionally controlled refrigerated
circulators.
out_hil_01
xxx
Set the desired maximum heating power (10 % to 100 %).
out_par_04
xxx
Band limiting during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.
out_par_06
xxx
Xp control parameter of the internal controller. (**
out_par_07
xxx
Tn control parameter of the internal controller. (**
out_par_08
xxx
Tv control parameter of the internal controller. (**
out_par_09
xxx
Xp control parameter of the cascade controller.
(**
out_par_10
xxx
Proportional portion of the cascade controller. (**
out_par_11
xxx
Tn control parameter of the cascade controller.
(**
out_par_12
xxx
Tv control parameter of the cascade controller. (**
out_par_13
xxx.xx
Maximum internal temperature in case of cascade control
out_par_14
xxx.xx
Minimum internal temperature in case of cascade control