Safety sensor in the cooling zone: The temperature in the cooling zone has
exceeded 170 °C. See „Important notice“ Page 23.
Press enter
to quit the audible signal
.
After eliminating the malfunction, press the mains power switch off and on again
to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized JULABO
service station.
Mains Fuses
:
Mains circuit breakers (resettable) for the circulator (16 A) and
the CU-unit (4 A) on the rear of the housing are safety machines.
Mains Fuses
:
The mains fuses (36) on the rear of the unit may easily be exchanged as
shown on the left – T 16 A.
Warning:
Before exchanging the fuses, turn off the mains power switch and disconnect the power plug
from the mains socket!
Only use fine fuses with a nominal value as specified.
Example:
Manufacturer
Supplier
Type
Order No.
Schurter
Schurter
G-fuse insert SPT
T16A 5x20mm
No. 0001.2516