Menu functions
42
8.3. Start of a profile
1.
2.
The start menu of the High Temperature Circulator allows calling up and
defined starting of one of six previously stored temperature profiles. This
start can be effected manually or be released by an integrated timer.
Important:
In the menu configuration under menu point „setpoint“ operation has to
be set via „key“.
There are two possibilities for manually starting a program:
1. Starting a program from the OFF status:
The programmer switches back to the OFF status at the end of the
program.
2. Starting a program from the operating status.
The programmer is started with the Start key
, and the bath is
heated to the desired temperature, for example 100 °C. At the end of
the program, the programmer switches to the operating status and
holds the bath temperature stable at 100.00 °C.
Press enter
to select the submenu „Profile Start“.
Use the up/down cursor keys
to select the desired
option.
A flashing segment indicates that a number needs to be entered.
Start Profile
0 to 5
at Step
0 to 60
Loops
1 to 99
Enter the desired number and set each entry with enter
.
Start
no / yes
(
m a n u a l s t a r t
)
or
time
(via integrated timer)
A flashing line indicates that a parameter needs to be entered.
Press the P-key
to select the respective parameter and press
enter
.