Menu functions
36
8.1. Configuration
By means of the configuration functions, operation of the High
Temperature Circulator can be optimized for the current application.
Press enter
to select the configuration submenu.
Use the up/down cursor keys
to select the desired
option. A flashing line indicates that a value needs to be entered.
Press the P-key
to select the parameter and press enter
.
Press escape
to return the previous menu level.
Identification
When performing an identification for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv will be
automatically determined and stored.
Possible parameters:
off
- no identification.
The control parameters ascertained during the last identification are
used for control purposes.
once
- single identification
The High Temperature Circulator performs a single identification of
the controlled system after start.
After the identification process the parameter is automatically set to
“off“.
always
- continual identification
The High Temperature Circulator performs an identification of the
controlled system whenever a new setpoint is to be reached.
NOTE:
Use this setting only when the temperature application
system changes permanently.
Note:
Requirement for an identification of the controlled system:
The High Temperature Circulator must heat to a setpoint temperature
at least 50 K above the previous setpoint using the adjusted heating
power.
For large external consumers, the setpoint span can be
selected even larger (100 K).