Menu functions
52
8.7. Interface
The interface parameters are set by selecting the submenu „Interface“
on the programmable controller. Normally, this is a one-time-only
adjustment.
Press enter
to select the submenu „Interface“.
Select the desired option with the up/down cursor keys
.
Enter the desired value for the flashing digit.
Type
RS232 / RS485
Baudrate
2400/4800/9600
Parity
none/even/odd
Handshake
software handshake/hardware handshake
Address
0 to 127
Press the P-key
to select the desired parameter and set with
enter
.
Return to the previous menu level with escape
.
Factory settings
see - 11.1.
8.8. Sensors
ATC - Absolute Temperature Calibration
Select the submenu „Temp.Sensor“ with enter
.
Select the desired option with the up/down cursor keys
.
A flashing digit indicates that a value needs to be entered i.e. set.
ATC Int:
internal sensor
ATC Ext:
external sensor
Maximum setting ±9.99 K
Enter the desired compensating value and set this value by pressing
ENTER