Safety notes for the user
18
Notice:
Check the safety installations at least twice a year!
Excess temperature protection according to IEC 61010-2-010
With a screw driver turn back the adjustable excess temperature protection until the shut-
down point (actual temperature).
Low level protection
To check the function of the float of this unit it cannot be operated manually.
The 5-graded level display should therefore be observed whenver refilling.
If the bath liquid thickens or cracks, the instrument should be cleaned and checked by
qualified personnel.
Caution:
Escape of vapors / gas
The necessity of degassing requires that a closed system is not entirely sealed. Especially at
increased working temperatures vapor / gas may escape.
Ensure sufficient ventilation at the place of installation!