Performing an Exam
Prior to beginning a test, ensure that the test defaults are set as desired (i.e. unit of measurement, number of
tests, number of repetitions, etc.). Once test defaults are set, choose the “Test” option from the main menu,
and follow the instructions listed under the specific instrument you will be testing with.
“Wireless Dual Inclinometers (9RF303, 9RF304)” on page 20.
“Wireless Muscle Tester (9RF305)” on page 24.
“Wireless Grip (9RF306)” on page 26.
“Wireless Algometer (9RF307)” on page 28.
“Wireless Goniometer (9RF308)” on page 30.
“Wireless Pinch Gauge (9RF309)” on page 32.
“Wireless Static Force Gauge (9RF310)” on page 34.
Reviewing and Deleting a Test
Once a test has been completed, it will be stored on the Commander Echo Console until it is deleted. To view
the completed test data, follow these steps:
1 .
From the main menu, select “Review.”
2 .
Move the Control Stick up or down to navigate between the completed tests.
3 .
Press the Control Stick to view more specific information regarding a specific test
(Note: moving the Control Stick up and down while viewing specific test information will move between
tests).
4 .
Move the Control Stick left and right, and press to view more information for a specific side, or to exit the
test information screen.
5 .
Hold down the Control Stick to delete individual tests.
Calibration Settings
Restoring Factory Calibration
Each instrument has been factory calibrated to ensure accuracy. It is possible that the user calibration may be
different from these pre-set parameters, which may lead to invalid force/angle reading. Prior to performing a
full calibration, attempt to restore the factory calibration stored on each instrument. If this does not correct the
problem a full calibration may be required.
Follow these steps to restore the factory calibration:
1 .
Select “Settings” from the main menu.
2 .
Select “Instrument” from the Settings menu.
3 .
Select the Instrument, for which you wish to restore the factory calibration.
4 .
Highlight the “System” option in the Instrument menu.
5 .
Move the Control Stick left or right until you highlight “Restore.”
6 .
Press the Control Stick to select restore.
7 .
The LCD screen will display the instrument ID at the top and Disconnected at the bottom.
8 .
Press the “On” button for the corresponding instrument followed by pressing the Control Stick on the
Console to establish connection.
9 .
The LCD Screen will show the device is connecting.
10 .
If the instrument connects to the Console, the LCD screen will display “RESTORE CALIBRATION.”
i .
Select, “No” to return to the Settings menu.
ii .
Select “Yes” to restore the factory calibration settings.
Note: restoring the factory calibration only overwrites the user calibration data with the factory calibration
data stored on the instrument.