© 2017 JTECH Medical MN022_M
8
Warnings and Notifications
Protection against ingress of liquids: Not protected against ingress of liquids. Keep Dry, do
not immerse in water or autoclave any portion of Commander Echo Console, instruments, or
accessories.
Type of protection against electric shock: Type B applied part.
Devices emit non-ionizing radiation.
Warning:
Devices are not sterile.
Warning:
Echo devices should only be used by trained professionals.
Contraindications
Warning:
This device should not be used in the following circumstances:
•
On or near the eye
•
On or near bone fractures
•
On or near open wounds
•
On or near burned tissue
•
With patients suffering with severe osteoporosis
•
For any purpose other than indicated
Warning:
No modification of equipment is allowed. Do not open the Console, instrument,
transceiver or receiver housings. Opening of housings by anyone other than an authorized JTECH
service representative will void your warranty.
Warning:
Changes, or modifications, not expressly approved by JTECH Medical could void the
user’s authority to operate the equipment.
Warning:
Use only a factory-supplied, medical-grade power supply certified to charge medical
devices. Use of another charger may result in electrical shock or equipment damage. It is not
recommended that you charge the devices by connecting them to a computer.
Warning:
Echo instruments are not intended for use while attached to the charger. Never
attempt to operate the instrument while it is connected to the charger as electrical shock or
damage to the instrument may occur.
Warning:
Echo instruments have no user serviceable parts.
Warning:
This device is not suitable for use in the presence of flammable anesthetic mixture with
air, or with oxygen or nitrous oxide
Warning:
Follow all directions as specified by the CaviWipe product being used. Allow devices to
dry before use.
Notice:
Echo instruments are precision medical devices and should be treated with care. Avoid
dropping, banging, or other impacts to the instruments.
Notice:
Use only within designated operating parameters, i.e. temperature and humidity limits.
Warning:
Discontinue use of any product if skin irritation develops.