DINAMAP*
Compact
Vital Signs Monitor
13
POWER CONVERTER:
Protection against electrical shock - Class 1
INPUT VOLTAGE:
115 / 230 VAC, 50 / 60 Hz (nominal),
90 ~ 264 VAC, 47 ~ 63 Hz (range)
MONITOR:
Protection against electrical shock - Class 1
INPUT VOLTAGE:
24 VDC (nominal), 12-30 VDC from supplied power converter
EXTERNAL DC LINE FUSE
: T3.15A 250 VAC
BATTERY:
12 volt, 2.3 amp-hours.
Minimum operation time: 2 hours (5 minute auto cycle with
adult cuff at 25°C with power save mode enabled) from full
charge. Time for full recharge: 8 hours from full discharge
Environmental
OPERATING TEMPERATURE:
+ 5° C to + 40° C (+ 41° F to + 104° F)
OPERATING ATMOSPHERIC
700 to 1060 hectoPascal
PRESSURE RANGE:
STORAGE TEMPERATURE:
– 20° C to + 50° C (– 4° F to + 122° F)
STORAGE / TRANSPORTATION
500 to 1060 hectoPascal
ATMOSPHERIC PRESSURE :
HUMIDITY RANGE:
0 % to 95 % non-condensing
RADIO FREQUENCY:
Complies with IEC Publication 801-3 (Draft 7, second edition)
ELECTROMAGNETIC IMMUNITY:
3.0V/m, modulation 80% AM @ 1kHz, frequency range
26.0~1000.0Mhz
This product conforms with the essential requirements of the
Medical Device Directive. Accessories without the CE Mark are not
guaranteed to meet the Essential requirements of the Medical
Device Directive.
The DINAMAP* Compact monitor is protected against vertically
falling drops of water and conforms with the IEC 529 standard at
level of IPX1. No harmful effects will come of vertically falling
drops of water making contact with the monitor.
0086
IPX1