46
Shaking trouble
①
If there is extremely intense vibration of the machine than usual
during operation, check the horizontality. If you need, please adjust
horizontality according 3.4.2
②
If there is extremely intense vibration of the machine and shaking
speed is slow down, check the sample location and relocate the
samples harmoniously
③
In case of overloading, control the sample weight according to
9.1(Max load per speed).
④
Check door switch if shaking is not operated after shaking start
button press.
⑤
Check Timer close status if shaking is not operated after shaking
start button press, cancel timer set-up.
⑥
If the problem is not fixed with the above method, please request
A/S.
Noise
①
Check fixed condition of platform or sample.
②
Check the weight of sample is shifted to one side, relocate the
samples evenly.
③
If the problem is not fixed with the above method, please request
A/S.