33
4.4.2 External Start/Stop Key
External Start/Stop button is on the control penal as membrane button. It generates
start and stop. Set the applying range of it.
1
Only Shaker
The button only works for shaker’s start/stop.
2
Only Incubator
The button only works for incubator’s start/stop.
3
Shaker Incubator
The button only works for shaker and incubator’s start/stop.
Shaker and incubator are both available for the external membrane button. Once one of the
shaker/incubator is on the generating and touch the button, the stopped part is operated. Touch
the button again, both of mode are stopped.
4.4.3 Calibration
The temperature sensor and rpm reached to the user’s reference measure value, accurate experiment
circumstance can be made. Calibration for coordinate is available.
4.4.3.1 Temperature Calibration
Calibration for temperature sensor of main controller. The unit provides three points calibration.
Each point of three shows the uncalibrated and calibrated status. Touch the point you want to
change. Touch the one point of Calibration 1,2,3. Put the external sensor on the chamber and close
the door.