4.4.3.3 Touch Screen Calibration
Touch penal coordinate could be in the problem for the reason of the older unit and the problem in
the unit. The unit has a touch screen coordinate function. Touch the “+” on the screen three time s.
And repeat it two times.
4.4.4 Auto Tuning
Update the parameter of PID temperature control. Optimum temperature for the main using
temperature is available. The unit is set to use the temperature on proper range. Once the
temperature is not actual on the main points during the actual use on the field, operating auto
tuning will tune the temperature parameter and make the unit operates the temperature on the
actual temperature.
Touch the Set Temp. and put the desired value. Touch the START button and start the parameter
tuning. Touch the stop button during the tuning and stop the tuning. Once the tuning is complicated,
touch the ok button and save it.
4.4.5 Operating Parameters
Set the parameters for the other operating.
1
Position Sensor
Check the platform stop on the right position.
2
Stall Check
Check the belt when the platform is not working properly.
3
Defrost Use
Defrost is not for the IST model. Off the defrost use.
4
Load Unbalance
Once the vibration is generated, the shaking speed is
adjustable.
5
Threshold
Load Unbalance check the vibration on the unit and adjust the
rpm automatically. The vibration is on the range of not
controlling, the automatic adjusting is operated.
Summary of Contents for IST-3075
Page 12: ...6 Foot crush Hand crush or pinch Lifting hazard Do not take the device apart deliberately ...
Page 19: ...13 ...
Page 22: ...16 ...