26
4.2.2 Shaking Control
(1) Start shaking control
Touch the Set RPM Button under the shaking stop status. Setting mode is appeared. Set the value on
the display and touch the Start button. Shaking control starts. SV/PV rpm is changed to green color.
(2) Stop the shaking control.
Touch the Set RPM button under the shaking operating status. Setting mode is appeared. Touch the
Stop button. Shaking stops. SV/PV rpm is changed to gray color.
(3) Change the shaking speed
Touch Set RPM button. Setting mode is appeared.
Set the value on the display and touch the Enter button. The rpm is changed.
Move out from the setting mode with ESC button.
In the case of no touch for 10 seconds, the display changed to the main automatically.
Check the PV, RPM on the setting display is available. (right upper side)
The operating is maintained as previous during the setting.
Once the door is opened during the shaking operating, the heater and fan is stopped. Close the
door before door error generating, the previous operating is back.
In the case of change the sample, stop the shaking and check whether the platform is
completely stopped.
Summary of Contents for IST-3075