28
Timer setting is available for hour:minute. Touch the part you want to change and put the
numbers on it.
Move out from the setting mode with ESC button.
In the case of no touch for 10 seconds, the display changed to the main automatically.
On the timer set display, you can check the present chamber temperature and rpm. (right upper
side of the display)
Below 1 minutes is displayed as“mmm:ss”.
Timer is not counted during the door open.
4.2.4 Alarm and Stop by force
The unit senses the problem on the unit and warn to the user. The warning is visual-audial effect.
The alarm is divided into warning and fault.
4.2.4.1 Warnings
Warning generates visual-audial alarm but the unit is operating. Once the warning is generated,
Silent button is appeared on the main display. Once the silent button is touched, next 15 minutes
there is no alarm. If the problem is on-going after 15 minutes, the alarm is generated again.