35
Once the door is opened or the temperature is out of range, the temperature control is
stopped.
The calibration is operating on various points so it limits the calibrating that is our of the logic.
The result of the calibration is on the uncalibrated and calibrated. Uncalibrated means the value
from the unit and the calibrated means the temp that calibrated. The calibrated value is from
the external sensor.
The calibration setting is available to be initialized on the System Setting Reset.
4.4.3.2 Shaking Speed Calibration
The user calibrates the unit’s rpm with external sensor. The unit is available for one point rpm
calibration. The procedure is similar with temperature calibration. The difference is that the shaking is
not stopped during the door opening because the door need to be opened to check the shaking
with external sensor. Touch the set rpm. Put the desired rpm and start the nut. Wait unit the actual
rpm reached to the set rpm. Touch the calibrated rpm and put the value from the external sensor.
Touch the confirm button to save it. The unit read the calibrated value from the next operating.
Be careful that the external sensor never touch the platform. Once the external sensor is
removed from the platform during the operating, it will be dangerous.
The calibration setting is available to be initialized on the System Setting Reset.