Chapter 10
Test Results
Copper results
OneExpert DSL User’s Guide
April 2014
22052280, Revision 001
89
Opens/Capacitance
You measure opens/capacitance to:
•
measure total electrical loop length (includes length of bridged taps).
•
find wet sections, bridged taps or open faults.
Opens/Capacitance: < 3% difference between Tip (A) and Ring (B) values indicate an acceptably
balanced loop. Anything greater indicates an unbalanced line or open fault, which inhibit digital
signals.
describes the opens/capacitance result.
DC current
You measure DC current to verify if enough current is present to operate equipment including the
phone or caller ID boxes.
DC current on Tip (A) to Ring (B) should be
23 mA at the NID. Anything less will not allow for differ-
ences in temperature and phone equipment will not operate during extremes.
describes the DC current result.
50 to 999
1
5% ± 2.5
1.0K to 9.99K
10
5%
10.0K to 99.9K
100
5%
100.0K to 999K
1K
5%
1.0M to 9.9M
10K
10%
10.0M to 100M
100K
15%
Table 24
Leakage result specifications (no ambient) (Continued)
Range (
)
Resolution (
)
Accuracy
Table 25
Opens/Capacitance result specifications (no ambient)
Range (feet)
Resolution
(feet)
Range (meters)
Resolution
(meters)
Range
(capacitance)
Accuracy (applies to
capacitance only)
0 to 2,999
1 foot
0 to 999
0.1 meter
0 to 47.1 nF
1% ± 15 pF
3000 to 66000
1 foot
1,000 to 20,000
0.1 meter
47.1 nF to 1.57 µF
2%
Table 26
DC current result specifications
Range
Resolution
Accuracy
0 to 110 ma
1ma
2% ± 1ma