R-HAx
W-LAx
R-LAx
Read - Higher
Temp.
Alarm + port
Write - Lower
Temp
Alarm + port
Read - Lower
Temp
Alarm + port
To get the Upper Temperature Alarm limit.
The ‘x’ must be replaced with the number of the
destination port in ACCII, for example “RHA1”.
The JBC device answers with an “A-HAx”.
The Data Field contains the upper Temperature Alarm
limit, expressed in °C.
Sets the lower Temperature Alarm limit.
The ‘x’ must be replaced with the number of the
destination port in ASCII, for example, “WLA1”.
The Data Field contains the Lower Temperature Alarm
limit, expressed in °C. The temperature modification
must be done in steps of 5 degrees.
The JBC device responds with an “A-LAx”.
To get the Lower Temperature Alarm limit.
Replace the ‘x’ with the number of the destination
port in ASCII, for example, “RLA1”.
The JBC device responds with an “A-LAx”.
The Data Field contains the lower Temperature Alarm
limit, expressed in °C.
Code
Description
Details
18