Predrying holding time Stage 2
This parameter defines the duration of the predrying process for Stage 2 once the desired predrying
temperature has been reached.
Value range OFF or 00:00 – 60:00 (mm:sss)
Thermo Shock Protection (TSP)
The TSP function protects the dental-lab work during the closing process. For this purpose, the TSP function
gauges the temperature of the firing chamber in the furnace head upon the start of the firing program. If
required, the closing path within the set closing time S is adjusted.
Setting options: On / Off
Pre-vacuum
If a firing program with pre-vacuum is conducted, the vacuum pump is switched on at the end of the
closing time (as soon as the furnace head is closed) and runs until the pre-vacuum time has elapsed. The
heating phase begins after the pre-vacuum time is over.
The value V1 is ignored upon the start of a program with individually activated pre-vacuum. The vacuum is
maintained until V2 is reached. V2 must be higher than the stand-by temperature B.
Setting options: OFF or 01:00 – 05:00 (mm:ss)
Hv
Vacuum holding time
With this function, the vacuum share of the holding time can be individually set.
Example: H (Holding time) = 02:00 (mm:ss). If a vacuum share of 50% is desired, the parameter "Vacuum
holding time (Hv)" has to be set to 01:00 (mm:ss).
Setting options: OFF or 00:01 – 60:00 (mm:ss)
Quick opening of the furnace head
If the option "Quick opening of the furnace head" is activated, the furnace head opens at maximum
speed at the end of the holding time.
Setting options: On / Off
Extraction system
If this function is activated, any vapours are extracted during the closing process by means of the vacuum
pump.
Setting options: On / Off
IoT function
The IoT function can be activated and deactivated here. Moreover, the connectivity status is indicated.
In addition, the time zone can be set in this section.