5. Operation and Configuration

4.  The write-protection was successfully activated.

 The lock 

symbol is shown next to the 

[Options]

 button in the display.

5.4.4  Adjustable parameters in the program screen 

t – Temperature increase rate  

The temperature increase rate defines by how many degrees per minute the furnace heats up.

Value range °C: 10 – 140 °C/min; Value range °F: 18 – 252 °F/min

T – Holding temperature 

The holding temperature defines the temperature at which the press process is started.

Value range °C: 100 – 1200 °C; Value range °F: 212 –  2192 °F

H – Holding time 

The press process starts once the holding time ends.

Value range: 00:00 – 60:00 (mm:ss)

E – Stop speed

This parameter defines the end of the press process.

For the layering technique, Ivoclar Vivadent recommends a stop speed of 300 µm/min and for the staining 
technique a stop speed of 150 µm/min.
–  Higher values (stop speed of e.g. 300 µm/min) result in the press cycle being stopped sooner.
–  Lower values (stop speed of e.g. 100 µm/min) result in the press cycle being stopped later. Consequently, 

the press cycle is prolonged.

Value range: 0 - 10,000 (µm/min)

5.4.5  Adjustable parameters in the Options menu

In addition to the parameters shown in the firing curve, several other options are available that can be activated by 
pressing the 

[Options]

 button. Active options are displayed as icons in the grid next to the 

[Options] 

button.

The following press program options are available:

Program write-protection

If the program write-protection is activated, the program parameters and program options cannot be 
changed. This is to prevent accidental changes to the program. 

Setting options: On / Off

  B

Stand-by temperature 

The stand-by temperature is the temperature to which the furnace heats up immediately after switching 
on. The temperature is maintained with the furnace head closed and when no firing or press process is 
activated.

In the "Pressing" mode, Programat furnaces feature an automatic stand-by temperature of 700 °C as the 
factory setting. The temperature can be individually set for each program.

Value range °C: 100 – 700 °C; Value range °F: 212 – 1292 °F

Summary of Contents for 645983EU

Page 1: ...Operating Instructions Valid from software V7 0 Programat EP 3010 G2 ...

Page 2: ...2 Declaration of conformity ...

Page 3: ...rograms and programming options 42 5 5 Advanced functions of the furnace Software update 46 5 6 IoT function 58 6 Practical Use 59 6 1 Firing procedure with an Ivoclar Vivadent program 59 6 2 Firing procedure with an individual program 60 6 3 Press procedure using an Ivoclar Vivadent press program 62 6 4 Press procedure by means of FPF fully automatic press function 63 6 5 Press procedure using an...

Page 4: ...5 Keypad 6 Optical Status Display OSD A Press drive cover 7 Cooling tray B Ventilation grid press drive 11 Covering hood for head opening mechanism 12 Air vents furnace base 13 USB connection 14 USB device 15 Ethernet connection 8 Screw for cooling tray 9 Furnace base housing 10 Furnace feet ...

Page 5: ...ng mechanism 21 QTK2 heating muffle 22 Sealing rims of furnace head 23 Sealing surface 33 Screw for connection cover C Screw for press drive cover 34 Air vents rear panel 35 Heating element fuse 36 Vacuum hose connection 37 Vacuum hose 29 Vacuum pump socket 30 Power cord for vacuum pump 20 Speaker 24 On Off switch 25 Power socket 26 Power cord List of parts ...

Page 6: ...63 Split taper socket for press plunger 64 Terminal screw for press plunger 65 Press plunger 66 Head insulation 61 Press electronics 38 Furnace head mounting 39 Furnace head release 44 Heater plug 45 Heater plug socket List of parts 67 LAN download cable 68 Programat Firing Tray Kit 2 69 Programat USB stick 70 Cooling grid 42 Furnace head vacuum hose 51 Press drive cable 41 Thermocouple cable ...

Page 7: ...erating Instructions The signs and symbols used in these Operating Instructions facilitate the finding of important points and have the following meanings Symbol Note Risks and dangers Important information Contraindication Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions Furnace concerned Programat EP 3010 G2 Target group Dental t...

Page 8: ...voltage range shown on the images e g rating plate may differ depending on the voltage version of your furnace 1 5 Notes on the images in these Operating Instructions All images and illustrations in these Operating Instructions are used for exemplification and the details are not authoritative for the construction of the furnace They are symbols which may slightly differ from the original e g due ...

Page 9: ...nstructions to assure proper use of the system The instructions regulations and notes in these Operating Instructions must be observed The instructions regulations and notes in the material s Instructions for Use must be observed The system must be operated under the indicated environmental and operating conditions Chapter 9 3 The Programat EP 3010 G2 must be properly maintained Risks and dangers ...

Page 10: ...n the firing chamber is damaged There is a risk of electric shock upon contact with the heating wire Avoid damage of the insulation by contact with the investment tongs or firing tongs Contraindications Firing trays must not be placed in the area surrounding the firing plate since this will obstruct the closing of the furnace head Foreign objects must not be placed on the furnace head or the air v...

Page 11: ...k This product contains ceramic fibres and may release fibre dust Do not use compressed air thus distributing the dust in the environment and observe the additional notes on page 13 Risk of crushing burn hazard Never reach under the furnace head with the hand or other parts of the body There is a risk of burning and crushing Never reach inside the rear cover with the hand or particularly with the ...

Page 12: ...y cloth Do not use solvents Disconnect the power plug before cleaning and allow the furnace to cool down The furnace must be cool before it is packed for transportation Use original packaging for transportation purposes Before calibration maintenance repair or change of parts the power must be disconnected and the furnace has to be cool if it has to be opened If calibration maintenance or repair h...

Page 13: ...mic fibres and may release fibre dust Fibre dust has proved to be carcinogenic in animal experiments The heating muffle may only be dismounted by a qualified After Sales Service Center Information regarding the Safety Data Sheet is also available from your After Sales Service Center Disposal The furnace must not be disposed with the common domestic waste Please correctly dispose of old furnaces ac...

Page 14: ...es are continuously compared The Programat EP 3010 G2 consists of the following components furnace base with electronic controls furnace head with firing chamber and press drive cooling tray firing plate power cord and hose for vacuum pump 3 2 Hazardous areas and safety equipment Description of the risk areas of the furnace Hazardous area Type of risk Firing chamber Risk of burning Opening closing...

Page 15: ...r missing contact your local Ivoclar Vivadent Service Center We recommend keeping the original packaging for future service and transportation purposes 4 2 Selecting the location Place the furnace on a flat table using the rubber feet Make sure that the furnace is not placed in the immediate vicinity of heaters or other sources of heat Make sure that air may properly circulate between the wall and...

Page 16: ...plate holder Place the cooling tray 7 on the frame plate 19 Make sure that the cooling tray is correctly positioned on the frame plate 4 3 Assembly Assembling the furnace is very easy and involves only few steps Before you start assembling the furnace make sure that the voltage indicated on the rating plate 27 complies with the local power supply If this is not the case the furnace must not be con...

Page 17: ...e furnace head with the furnace base Proceed as follows Connect the vacuum hose Insert the thermocouple plug make sure that the polarity of the plug is correct Insert the press drive plug Insert the heater plug 1 Insert the heater plug into the intended connector 2 Secure the heater plug by rotating it 45 until it snaps into place Position the furnace head mounting as shown in the picture until th...

Page 18: ...ially coordinated with the press furnace If other pumps are used please observe and do not exceed the maximum power consumption Do not shorten the vacuum hose The minimum length of the vacuum hose is 1 6 m Step 6 Mount the connection cover Mount the connection cover and secure it with the fastening screw Make sure that the lateral catch springs snap into place when mounting the connection cover Th...

Page 19: ...over 3 Open the covering hood 4 Release the heater plug and disconnect it with a 45 anti clock wise rotation 5 Disconnect the press drive plug 6 Disconnect the thermocouple plug 7 Disconnect the vacuum hose 8 Press the leaf spring with a finger lift off the furnace head at the same time and remove it Make sure that the furnace head has completely cooled down before it is removed fire hazard 4 5 1 ...

Page 20: ...pon the initial start up of the new furnace a number of basic settings are required These settings will be stored and will not appear anymore upon the following starting procedure Select the desired language using the Arrow up down buttons Confirm the entry with the green button The Next button is used to reach the next entry screen Step 1 Step 2 Step 3 Step 1 Select the language The first setting...

Page 21: ...t entry screen Step 6 Set the date day month year Confirm the entry with the green button The Next button is used to reach the next entry screen The initial start up and entry of the basic settings are now complete The furnace will now automatically conduct a self test Step 7 Confirm the privacy policy More information about the privacy policy can be found at http www ivoclarvivadent com digitalse...

Page 22: ...s in the acceptable range The power supply is outside the acceptable range Software version The currently installed software version is displayed If the self test has been successful the furnace will automatically display the screen to select the desired operating mode firing pressing If the program recognizes a malfunction during the test a corresponding error message with the corresponding recti...

Page 23: ...ving function is activated only possible with the furnace head closed and the furnace on idle The display shows the power saving icon Pressing any key ends the power saving function OPEN FURNACE HEAD key Quick cooling with the furnace head open If the furnace head is completely open and the OPEN FURNACE HEAD key is pressed again the quick cooling function is activated e g the vacuum pump is switch...

Page 24: ...u level The button indicates to which screen you change e g to the home screen Close With this button you can leave sub menus You change to the next higher menu level Confirm entry This button is used to confirm an entry If the button appears pale green no entry has been made yet or the value entered is not in the acceptable range Cancel entry This button is used to cancel an entry Changed values ...

Page 25: ...ed button without any parameters being changed 5 1 5 Selecting the operating mode The furnace can be operated in two different modes Pressing Only press programs can be selected The stand by temperature of the furnace is set to 700 C Firing Only firing programs can be selected For Ivoclar Vivadent programs the stand by temperature of the furnace is set to 403 C In order to switch between the press...

Page 26: ...his operating mode only firing programs are available and the stand by temperature is set to 403 C 5 1 6 Explanation of the home screen After the furnace is switched on the display shows the home screen All functions of the Programat can be selected from this screen You can switch back to the home screen with the HOME key By pressing a selection button you will reach the corresponding menu e g fir...

Page 27: ...s is acknowledged with the STOP key the signal transmitter is switched off and no further signals will be sounded For error messages Error messages are acoustically supported with the error melody endless beep The acoustic signal can be confirmed by pressing the STOP key while the error message still remains visible If the error message is acknowledged with the corresponding button the signal tran...

Page 28: ...if the programs are to be used for other purposes Therefore these programs are also available as individual programs In case of software updates individually changed parameters in Ivoclar Vivadent programs might be reset to the factory settings or changed b 300 free individually adjustable programs The programs are designed in such a way that they can be either used as conventional one stage progr...

Page 29: ... 2 Select the program group Quick program selection Pressing the P key switches to the program screen of the current program When the program screen is shown pressing the P key again switches to the quick program selection by means of number entry 5 Operation and Configuration 5 2 2 Program selection The program selection requires only few steps 1 Select the program type ...

Page 30: ...ed or edited in this screen For Ivoclar Vivadent programs the write protection must be deactivated first before any parameters can be changed The following information is displayed 1 Information bar Program designation Current furnace temperature 2 Firing curve Closing time holding time Temperature increase rate holding temperature long term cooling Vacuum on vacuum off 3 Program options In additi...

Page 31: ...ctivating the write protection 1 Press the Options button 4 The write protection was successfully deactivated The lock symbol is no longer shown in the display next to the Options button 2 Press the Deactivate Write Protection button 3 Press the Close button to leave the Options menu 5 Operation and Configuration ...

Page 32: ...the Predrying Temperature button 4 Enter the desired predrying temperature and confirm by pressing the green button 5 The predrying temperature was successfully changed Press the Close button to leave the Options menu The program screen now shows the symbol Predrying active next to the Options button 5 Operation and Configuration ...

Page 33: ...n Example 1 Changing the program stage The program screen shows the parameters of the first temperature level S H1 etc Pressing the Step button results in the screen for the second temperature level being displayed 2 Now the parameters for the second temperature level can be entered Pressing the Step button again switches the screen back to the first temperature level Automatic plausibility check ...

Page 34: ... long term cooling Value range C Off or 1 1200 C Value range F 0 or 34 2192 F Long term cooling If long term cooling is activated the furnace cools to the set temperature L at the end of the holding time with the furnace head closed Value range C Off or 50 1200 C Value range F 0 or 122 2192 F Cooling rate Can only be set if long term cooling L is activated This parameter defines by how many degree...

Page 35: ...perature The stand by temperature is the temperature to which the furnace heats up immediately after switching on The temperature is maintained with the furnace head closed and when no firing process is active Programat furnaces are programmed to a stand by temperature of 403 C in the factory The temperature can be individually set for each program Value range C 100 700 C Value range F 212 1292 F ...

Page 36: ...d upon the start of a program with individually activated pre vacuum The vacuum is maintained until V2 is reached V2 must be higher than the stand by temperature B Setting options OFF or 01 00 05 00 mm ss Hv Vacuum holding time With this function the vacuum share of the holding time can be individually set Example H Holding time 02 00 mm ss If a vacuum share of 50 is desired the parameter Vacuum h...

Page 37: ... the form of a firing curve The estimated remaining time is also displayed in 10 second intervals If no vacuum is activated the vacuum indicator and all the corresponding parameters are blank The process status is displayed in colour in the firing curve Orange Program is closing the furnace head or is in predrying mode Red Program is heating up or is in holding time mode Blue Program is in long te...

Page 38: ...play button is pressed while the firing curve is displayed the remaining time will be shown The remaining time indicator informs the user about the time that is left until the process is completed The remaining time is displayed in the centre of the screen in large characters and is thus visible even from a distance The factory settings of the furnace are such that the firing curve is automaticall...

Page 39: ... flashing Pause and the indicator changes back to the program screen If a program is stopped prematurely Vacuum release is shown in the display during the flooding of the firing chamber 5 2 8 Changing the parameters while the program is running Most program parameters which have not yet been executed can be changed while the program is paused Change the parameters as described in Chapter 5 2 3 5 O...

Page 40: ... want to copy 3 Select target The selection can be confirmed with the green button The red button is used to close the copy assistant 5 3 1 Copying programs With the copy assistant program sectors can be copied The source and target of the copy process can be selected in the copy assistant Programs copied to a USB flash drive can only be opened with this furnace PrograBase X10 offers the possibili...

Page 41: ...responding button in the Admin menu 2 Select the desired product logo Confirm the entry with the green button 5 3 3 Renaming program program group The current program and current program group can be renamed only possible for individual programs 1 Press the corresponding button in the Admin menu and select whether the program or group name should be changed 2 Enter the desired program or group nam...

Page 42: ...ess programs for Ivoclar Vivadent materials When the furnace is delivered ex factory the Ivoclar Vivadent programs already contain the recommended material parameters They cannot be deleted or overwritten The parameters are not shown in the software b FPF Fully automatic press function for IPS Press materials This function has been ideally coordinated with the IPS e max Press materials The ingot t...

Page 43: ...from Ivoclar Vivadent e g IPS e max IPS Empress Esthetic only the original Ivoclar Vivadent press programs must be used as they are especially coordinated with the respective materials After the individual press program was selected the following information is displayed on the program screen 1 Information bar Program name investment ring size Current furnace temperature 2 Firing curve Holding tim...

Page 44: ...he green button 2 Press the Activate Write Protection button 3 Press the Back button to leave the Options menu The holding temperature was successfully changed All the other parameters shown in the firing curve can be changed edited in the same way Changing the program options Pressing the Options button opens the menu for advanced program options Example 1 Activating the program write protection ...

Page 45: ...lues stop speed of e g 100 µm min result in the press cycle being stopped later Consequently the press cycle is prolonged Value range 0 10 000 µm min 5 4 5 Adjustable parameters in the Options menu In addition to the parameters shown in the firing curve several other options are available that can be activated by pressing the Options button Active options are displayed as icons in the grid next to...

Page 46: ...lect the desired display brightness Select the desired display brightness in percent and confirm the entry with the green button or cancel the entry with the red button The setting was changed To return to the home screen press either the Home touch button in the navigation bar or the HOME key on the membrane sealed keypad 5 5 Advanced functions of the furnace Software update 5 5 1 Settings To rea...

Page 47: ... Select the desired melody for the acoustic signals Setting options Melody 1 to 20 Time Set the current time Setting options hh mm ss Date Set the current date Setting options according to the set date format Date format Set the date format Setting options dd mm yyyy mm dd yyyy Time zone Set the respective time zone Setting options from 11 to 12 hours revolving Power saving mode If the power savin...

Page 48: ... the program the parameters used are logged and saved in the furnace If the furnace is connected to the PrograBase software the saved table entries are synchronized with the connected laptop PC Protocols can be edited saved and printed with the PrograBase software Setting options Inactive Printer PC Laboratory name Enter the name of the laboratory The name is automatically added to the protocols S...

Page 49: ... number of the furnace Software version Currently installed software version of the furnace Software updates are available from www ivoclarvivadent com downloadcenter Last calibration Date of the last calibration Mains voltage Currently measured mains voltage Last dehumidification Date of the last dehumidification of the furnace Operating hours Number of operating hours 5 Operation and Configurati...

Page 50: ...dress Press procedures Number of all press cycles Press cycles since last calibration Number of press cycles since the last calibration Video tutorials Display of the QR code to access the video tutorials 5 5 3 Temperature calibration The thermocouple and heating muffle of the furnace may be subject to changes which affect the furnace temperature depending on the mode of operation and the frequenc...

Page 51: ...led The calibration value is the difference between the measured current temperature and the desired set temperature At the end of the program open the furnace head and carefully remove the ATK2 sample using the furnace tongs and place it on the cooling tray to allow it to cool Reinsert the firing plate using the furnace tongs To return to the home screen press either the Home touch button in the ...

Page 52: ...vidual programs and settings can be backed up on a USB flash drive We recommend using this feature e g before a software update or before sending in the furnace for maintenance purposes Moreover furnace data saved on a USB flash drive can be restored on the furnace However restoring only works on the same furnace with the identical serial number The procedure is identical for both functions and re...

Page 53: ...dent programs may be overwritten during a software update Individual programs are not affected and will not be overwritten 5 5 5 Software update Software updates can be easily installed on the furnace by means of a USB stick A USB stick which contains the current software file e g EP 3010 G2_V1 10 iv is required for the update The software version on the USB stick must be higher than the one insta...

Page 54: ...hamber e g firing tray influences the result Conduct the heater test immediately after switching the furnace on and before any actual firing procedures are conducted If the furnace is too hot an incorrect heating muffle quality will be indicated If the quality of the heating element drops below 50 changing the heating element is strongly recommended Keypad touch test Each time the keypad or the to...

Page 55: ...ion data e g calibration values date of the last calibration etc test results and saved error messages Generating a diagnostic file 1 Open the diagnosis function Press the Remote Diagnostics button in the Diagnosis menu 2 Generate a diagnostic file Connect a USB stick with the furnace Then press the Execute button 3 Once the diagnostic file has been generated the following message is displayed Dia...

Page 56: ...n of the firing chamber or the vacuum pump results in a poor vacuum and thus in poor firing results For that reason the furnace head should be kept closed when the furnace is switched off or is below 100 C in order to prevent the absorption of humidity Conducting the dehumidification program 1 Open the dehumidification program Scroll to page 3 in the home screen and press the Maintenance Programs ...

Page 57: ...selected and started in this mode Programs cannot be changed Settings or special functions cannot be activated or changed The protected mode can only be activated or deactivated by entering the user code Production In this mode only one program can be accessed Only the functions OPEN CLOSE FURNACE HEAD and START STOP program are available Only that program is available which was loaded in the furn...

Page 58: ...n modes can be quit by prolonged pressing of the HOME key at least 3 seconds 5 Operation and Configuration 5 6 IoT function 5 6 1 Description of the IoT function The IoT function enables you to optimize your device software and supports the fast trouble shooting in case of a technical problem More information about the IoT function can be found at http www ivoclarvivadent com digitalservices ...

Page 59: ... button Select the desired program group The practical procedure of conducting a firing or a press cycle is explained by means of an Ivoclar Vivadent program and an individual program 6 1 Firing procedure with an Ivoclar Vivadent program Step 1 Switch the furnace on Switch the furnace on using the mains switch at the back of the furnace Subsequently the furnace conducts an automatic self test Wait...

Page 60: ...s Display lights up in green the furnace is ready for the next program start Step 2 Select the operating mode After the self test has been completed the selection of the operating modes is displayed Select the operating mode Firing Step 3 Select the program Press the Individual Programs button 6 Practical Use 6 2 Firing procedure with an individual program Step 1 Switch the furnace on Switch the f...

Page 61: ... Now open the furnace head using the OPEN FURNACE HEAD key and position the firing tray with the object in the furnace Step 6 Start the program Activate the program by pressing the START key The LED lights up in green The process is indicated in the firing curve display The furnace head opens automatically at the end of the program When the Optical Status Display lights up in green the furnace is ...

Page 62: ...g 300g Not all programs can be run with all the investment ring sizes Please also refer to the Instructions for Use of the respective material Step 3 Select the program Press the Ivoclar Vivadent Programs button Select the desired program 6 3 Press procedure using an Ivoclar Vivadent press program Step 1 Switch the furnace on Switch the furnace on using the mains switch at the back of the furnace ...

Page 63: ...lay lights up in green the furnace is ready for the next program start Step 2 Select the operating mode After the self test has been completed the selection of the operating modes is displayed Select the operating mode Pressing Step 3 Select the program Press the Fully Automatic Press Function 6 4 Press procedure by means of FPF fully automatic press function Step 1 Switch the furnace on Switch th...

Page 64: ...ogram will be started The LED lights up in green The progress of the program in shown in the display The furnace head opens automatically at the end of the program Press time and press distance are indicated in the results panel After acknowledging the results panel and as soon as the OSD lights up in green the furnace is ready for the next program start ...

Page 65: ...able for the individual press programs Step 3 Select the program After the self test has been completed the home screen appears on the display Press the Individual Programs button Select the desired program 6 5 Press procedure using an individual press program Step 1 Switch the furnace on Switch the furnace on using the mains switch at the back of the furnace Subsequently the furnace conducts an a...

Page 66: ...PEN FURNACE HEAD key and place the investment ring into the furnace Step 7 Start the program Activate the program by pressing the START key The LED lights up in green The progress of the program is shown in the display The furnace head opens automatically at the end of the program When the Optical Status Display lights up in green the furnace is ready for the next program start ...

Page 67: ...if the sealing rims of the furnace head and furnace base are clean and undamaged Sealing rims of the furnace head and base Weekly Check the keypad for visible damage If the keypad is damaged it has to be replaced by a certified Ivoclar Service Center Membrane sealed keypad Weekly Check the temperature Use the temperature checking set to check and adjust the tem perature in the furnace Firing chamb...

Page 68: ...rticularly if the temperature drops below 150 C If the furnace head is open there is a risk that the insulation absorbs humidity and condensation water develops during firing This negatively affects the vacuum performance and consequently the firing results 7 5 Power saving mode If the furnace is not used for longer periods of time Ivoclar Vivadent recommends to activate the power saving mode With...

Page 69: ... plunger the following procedure must be followed Step 1 Remove the attachment screw Remove the screw C while the furnace head is closed Step 2 Remove the press drive cover First firmly push the cover to the back arrow 1 and then remove it by lifting it arrow 2 Step 3 Loosen the press plunger Unscrew the clamping screw of the press plunger 64 by turning it approx one half turn ...

Page 70: ...h the other hand Contraindication Do not touch the thermocouple when replacing the press plunger Step 6 Push the white press plunger 65 with the taper ahead into the guide bush Push the press plunger with slightly rotating movements into its split taper socket 63 and fasten the screw 64 Note The furnace must only be operated with the press drive cover mounted Contraindication Never touch the press...

Page 71: ...rting the program later 16 T2 T1 Enter a lower value for T1 or a higher value for T2 17 Power failure 10 s during a program in progress A program in progress was interrupted for more than 10 s by a power failure The program cannot be continued 18 T1 V12 Enter a lower value for T1 or a higher value for V12 19 Yes V2 B Pre vacuum activated V2 value must be higher than B 20 No Error in the heating sy...

Page 72: ...quired final vacuum value cannot be reached Check the vacuum pump 801 Vacuum drop An unacceptable vacuum drop has occurred 802 The vacuum does not increase self test No vacuum increase could be measured Check the following points Is the firing chamber tight no contamination on the sealing surfaces Is the vacuum hose connected Is the vacuum pump connected Is the fuse F1 OK 803 The vacuum system is ...

Page 73: ... data backup Please check the individual program groups The individual program groups are no longer valid 1913 Yes Data backup recover The new software contains program parameters which were not yet available at the time of the data backup Please check the individual programs The individual programs are no longer valid 1920 Yes Data backup recover The data backup could not be performed Please inst...

Page 74: ...ram Is humidity condensation in the vacuum hose Start the dehumidification program Wrong or illogical temperature indication Is the thermocouple buckled or fractured Contact the Ivoclar Vivadent Service Center Is the plug for the thermocouple correctly inserted Insert the plug correctly Is the thermocouple plug defective Cracks in the insulation Are the cracks small and neglectable hairline cracks...

Page 75: ...g to factory settings In this process all programs melodies and volume adjustments are irrevocably reset to the factory settings For that purpose proceed as follows 1 Open the Settings menu Scroll to page 2 in the home screen and press the Settings button 4 Enter the user code Enter the user code 1234 and confirm the entry with the green button or cancel the entry with the red button 2 Open the se...

Page 76: ...The following message is displayed Reset to factory settings successful Reset to factory settings failed To return to the home screen press either the Home touch button in the navigation bar or the HOME key on the membrane sealed keypad 8 What If ...

Page 77: ...0 Max power consumption 12 A at 110 120 V 8 A at 200 240 V Acceptable data for pumps of other manufacturers Max output Final vacuum 250 W max leakage current 0 75 mA 50 mbar Only tested pumps should be used Values of electric fuses 110 120 V 125 V T15A heating circuit 125 V T5A vacuum pump 200 240 V 250 V T8A heating circuit 250 V T3 15A vacuum pump Dimensions of electric fuses 110 120 V Diameter ...

Page 78: ...mbient temperature 5 C to 40 C Acceptable humidity range Relative humidity 80 for temperatures up to 31 C gradually decreasing to 50 relative humidity at 40 C condensation excluded Acceptable ambient pressure The furnace is tested for use at altitudes of up to 2 000 m above sea level 9 4 Acceptable transportation and storage conditions Acceptable temperature range 20 C to 65 C Acceptable humidity ...

Page 79: ...e program tables can be downloaded from the Internet as PDF files Please note that the version of the firing program table should correspond to the software version which is used on your furnace 10 2 Press program table Material Investment ring size 100 g 200 g 300 g IPS e max Press Multi IPS e max Press HT IPS e max Press MT IPS e max Press LT IPS e max Press MO IPS e max Press HO IPS e max Press...

Page 80: ...adway BSD Block M5 No 1 Kecamatan Cisauk Kelurahan Sampora 15345 Tangerang Selatan Banten Indonesia Tel 62 21 3003 2932 Fax 62 21 3003 2934 www ivoclarvivadent com Ivoclar Vivadent s r l Via del Lavoro 47 40033 Casalecchio di Reno BO Italy Tel 39 051 6113555 Fax 39 051 6113565 www ivoclarvivadent it Ivoclar Vivadent K K 1 28 24 4F Hongo Bunkyo ku Tokyo 113 0033 Japan Tel 81 3 6801 1301 Fax 81 3 58...