3.
Place the thermometer(s) for calibration into a suitable insert(s) in the metal block and
wait for the temperature to stabilise, connect the standard thermometer to the indicator.
Ensure the indicator is configured for the correct sensor and where applicable the
calibration data has been entered and user calibration enable - see pages 16 to 19.
4.
When the temperature indicated by the controller and that of the other thermometers are
stable (see specification for typical values) record th re e sets of readings over a period
o f about six minutes. Check that these readings are consistent and use their ave ra g e
values for the final calibration figures.
5.
Compare the units under test to the value from the in-built indicator.
6.
If the Pegasus
Plus
1200 has been calibrated, correct the figures accordingly.
7.
Reset the controller and/or repeat the calibration for another thermometer.
8.
When the calibration is complete, reset the controller to 0 C and wait until the unit has
co o led to below 600 C before moving the Pegasus
Plus
1200 to a new location. T h e
Pegasus
Plus
1200 must be cooled to below 100 C before it can be put back into its
carrying case.
Summary of Contents for PEGASUS PLUS 1200 B
Page 3: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 3 of 50 ...
Page 14: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 14 of 50 ...
Page 17: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 17 of 50 ...
Page 25: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 25 of 50 Testing Thermostats ...
Page 44: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 44 of 50 FIGURE 1 ...