Err4
Error 4: Keyboard failure
Stuck button, or a button was pressed
during power up.
Switch the power off and then on
without touching any of the
controller buttons.
Err5
Error 5: Input circuit failure
Consult Isotech
Pwr.F
Power failure
. The line voltage is too
low
Check that the supply to the
controller is within the rated limits