PEGASUS
PLUS
1200
HANDBOOK
ISSUE 5 - 01/05
Page 26 of 50
The Site model can be used the Isotech Cal Notepad software for the testing of thermostats and
other thermal switch e s wi th vo l t free contacts. Cal Notepad can capture the temperature at
which a switch opens or closes. It can also perform a hysteresis test. Refer to the Cal Notepad
manual for details. Ensure only voltage free contacts are connected. Do not allow any voltage
signal to be connected - doing so many damage the equipment and present a safety hazard.
Summary of Contents for PEGASUS PLUS 1200 B
Page 3: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 3 of 50 ...
Page 14: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 14 of 50 ...
Page 17: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 17 of 50 ...
Page 25: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 25 of 50 Testing Thermostats ...
Page 44: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 44 of 50 FIGURE 1 ...