of about six minutes. Check that these readings are co n si stent and use their average
values for the final calibration figures.
Compare the units under test to the standard thermometer.
5.
Reset the controller and/or repeat the calibration for another thermometer.
6.
When the calibration is complete, reset the controller to 0 C and wait until the unit has
cooled to below 6 0 0 C b efore moving the Pegasus
Plus
1200 to a new location. The
Pegasus
Plus
1200 must be cooled below 100 C before it can be put back into its carrying
case.
PEGASUS
Plus
1200S
CALIBRATION USING A STANDARD THERMOMETER AND THE INTERNAL INDICATOR
1.
Remove the Pegasus
Plus
1200 from its case and visually inspect it for any damage it may
have sustained since it was last used. Insert the required metal insert into the furnace
block using the tool supplied to avoid damage to the heater assembly.
2.
Connect the Pegasus
Plus
1 2 00 to a suitable power supply and set the controller to the
required temperature.