PEGASUS
PLUS
1200
HANDBOOK
ISSUE 5 - 01/05
Page 39 of 50
6.
Reset the controller and/or repeat the calibration for another thermometer.
7.
When the calibration is complete, reset the controller to 0 C and wait until the unit has
cooled to below 600 C befo re m o vi ng the Pegasus
Plus
1200 to a new location. The
Pegasus
Plus
1200 must be cooled below 100 C before it can be put back into its carrying
case.
PEGASUS
Plus
1200B, 1200S
CALIBRATION USING A STANDARD THERMOMETER WITH EXTERNAL INDICATION.
1.
Remove the Pegasus
Plus
1200 from its case and visually inspect it for any damage it may
have sustained since it was last used. Insert the required metal insert into the furnace
block using the tool supplied to avoid damage to the heater assembly.
2.
Connect the Pegasus
Plus
1 2 0 0 to a suitable power supply and set the controller to the
required temperature.
3.
Place the thermometer(s) for calibration and the standard thermometer into suitable holes
in the metal insert; wait for the temperature to stabilise.
4.
When the temperature indicated by the controller and that of the other thermometers are
stable (see specification for typical va l u e s) record three sets of readings over a period
Summary of Contents for PEGASUS PLUS 1200 B
Page 3: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 3 of 50 ...
Page 14: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 14 of 50 ...
Page 17: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 17 of 50 ...
Page 25: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 25 of 50 Testing Thermostats ...
Page 44: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 44 of 50 FIGURE 1 ...