PEGASUS
PLUS
1200
HANDBOOK
ISSUE 5 - 01/05
Page 22 of 50
These values can be inspected, and modified with the following procedure,
Press the PAGE key until the display indicates,
ACCS
LiSt
Press the SCROLL key until the display shows,
Goto
OPEr
Press the UP key until the display shows
Goto
conF
Press the Scroll Key twice, when the display will show,
inSt
Conf
Press the Page Key until the controller shows
CAL
Conf
Now use the Scroll key to examine the data pairs. The values can be modified with the UP and
DOWN keys.
To exit this mode press the Page key until the top display shows,
Exit
And then set the lower display to YES . Wh i l e in this mode take care not to modify other
parameters - a full list of all the parameters can be found in appendix 3.
CALIBRATION DATA EXAMPLE
A maximum of five points may be entered, shown as Pnt 1 to Pnt 5 for the temperature point and
Summary of Contents for PEGASUS PLUS 1200 B