INITIAL TESTING
This unit was fully tested before despatch to you but please check its operation as outlined
below.
After connecting the Pegasus
Plus
1200 to the electricity supply, the temperature controller
display will show the temperature of the block and the last set-point value. The S controller
and indicator both go through a self-test sequence first. The fan on the front panel should be
heard running. Assemble the insert as shown on pages 13 and 14.
Change the set-point to 300 C and observe that the block temperature rises and settles to
this value. For the S; place a thermometer in an insert in the block and connect it to the
suitably configured indicator. Confirm that the indicator agrees within ±2 C of the controller.
Your unit should have performed as described above and can now be used for calibration.
If any problems or faults arise during these tests please contact us or our agents for help
and advice.
To prolong heater life we recommend that the temperature should not be
increased by increments greater than 500°C.
IMPORTANT NOTICE
The controller's function settings are preset and will not require adjustment.