Using the PC Interface
The
Plus
models include an RS422 PC interface, a special converter cable that allows use with
the a standard RS232 port. When using the bath with an RS232 p o rt i t i s e ssential that this
converter ca b l e i s used. Replacement cables are available from Isotech, part number
ISO-232-432. A further lead is available as an option, Part Number ISO-422-422 lead which
permits up to 5 instruments to be daisy chained together.
The benefit of this approach is that a number of calibration baths may be connected together in
a " d a i sy
c h a i n "
co n figurati
on - and then
linked to a
single RS232,
s
e
e
diagram.
Note: The RS 422 standard specifies a maximum lead length of 1200M (4000ft). A true RS422
port will be required to realise such lead lengths. The Isotech conversion leads are suitable for
maximum combined lead lengths of 10M that is adequate for most applications.
Connections
For RS232 use simply connect the Isotech cable, a 9 to 25 pin converter is included to suit PCs
with a 25 pin serial converter.
RS422 Connections
Pin
Connection
4
Tx+ A
5
Tx- B
8
Rx+ A
9
Rx- B
1
Common
Using the Interface
Summary of Contents for PEGASUS PLUS 1200 B
Page 3: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 3 of 50 ...
Page 14: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 14 of 50 ...
Page 17: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 17 of 50 ...
Page 25: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 25 of 50 Testing Thermostats ...
Page 44: ...PEGASUS PLUS 1200 HANDBOOK ISSUE 5 01 05 Page 44 of 50 FIGURE 1 ...