PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 46 of 142
88293-EN, Rev C
Style 1.0
A
<Titrate> - press to switch to single spot to perform test burns, adjustable up to 4
spots; press again to revert to prior laser mode
B
<Fixation> - press to turn on fixation light (when applicable)
C
<Outline> - press to enable full pattern outline (when applicable)
D
<Auto> - press to enable automatic advancement (when applicable)
E
<AIM> - press to turn on/off aiming beam in STANDBY (always on in READY)
Aiming beam intensity - scroll up/down to adjust
F
<Single Spot>/<Pattern> /<EpM> - press to display patterns available at default
parameter setting.
G
<Rotate> - press to rotate or advance pattern (when applicable)
H
<Center> - press to position the beam to optical center of the treatment
NOTE
•
Parameter controls only show on screens when applicable to the pattern selected.
See the section “Pattern Descriptions/Pattern Parameters” for detailed information.
•
Center Button requires the micro manipulator is not being used
B
G
E
F
A
C
D