PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 104 of 142
88293-EN, Rev C
Replacing the Fuses
To change the power receptacle fuses:
1. Ensure the key switch is in the OFF position.
2. Unplug the main power cable from the wall outlet and the system main power
receptacle.
3. Insert a small, insulated flathead screwdriver into the fuse holder release receptacle
and unlock and remove the fuse holder.
4. Replace the blown fuses with new fuses compatible with the mains supply voltage, as
indicated in the <System Specifications> section of this manual.
5. Replace the fuse holder.
Main power
cord receptacle
Fuse holder
Fuses
Location of
system fuses