PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 117 of 142
88293-EN, Rev C
Code
Message
Critical
Clearable
Warning
Description
Action
62
Red 60µm Treatment
Local Light Over (Low)
X
Laser output is 20% over the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
63
Red 60µm Treatment
Local Light Over
(Medium)
X
Laser output is 50% over the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
64
Red 60µm Treatment
Local Light Over
(High)
X
Laser output is 100% over the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
65
Red 200µm Treatment
Local Light Under
(Low)
X
Laser output is 20% under the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
66
Red 200µm Treatment
Local Light Over (Low)
X
Laser output is 20% over the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
67
Red 200µm Treatment
Local Light Over
(Medium)
X
Laser output is 50% over the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service