PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 110 of 142
88293-EN, Rev C
3D Controller not functioning.
Probable Cause:
Loose connection or disconnected from system.
Suggestion:
Ensure that cables are correctly connected. Reboot the
system.
Attempt to use 3D Controller when in “Standby” mode and
confirm parameters can be changed.
Trouble Adding/Removing Favorites.
Probable Cause:
Favorite’s Database Update Error.
Suggestion:
Reboot the System.