PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 114 of 142
88293-EN, Rev C
Code
Message
Critical
Clearable
Warning
Description
Action
29
Base Treatment
Temperature Low
(Severe)
X
The base treatment temperature is
severely low
Check fan and
ventilation. Allow system to
cool.
30
Base Treatment
Temperature High
(Moderate)
X
The base treatment temperature is
high
Check fan and
ventilation. Allow system to
cool.
31
Base Treatment
Temperature High
(Severe)
X
The base treatment temperature is
severely high
Check fan and
ventilation. Allow system to
cool.
32
BRF Treatment
Temperature Low
(Moderate)
X
Laser module treatment
temperature is low
Check fan and
ventilation. Allow system to
warm up.
33
BRF Treatment
Temperature Low
(Severe)
X
Laser module treatment
temperature is severely low
Check fan and
ventilation. Allow system to
warm up.
34
BRF Treatment
Temperature High
(Moderate)
X
Laser module treatment
temperature is high
Check fan and
ventilation. Allow system to
cool.
35
BRF Treatment
Temperature High
(Severe)
X
Laser module treatment
temperature is severely high
Check fan and
ventilation. Allow system to
cool.
36
LBO Treatment
Temperature Low
(Moderate)
X
Laser module treatment
temperature is low
Check fan and
ventilation. Allow system to
warm up.
37
LBO Treatment
Temperature Low
(Severe)
X
Laser module treatment
temperature is severely low
Check fan and
ventilation. Allow system to
warm up.
38
LBO Treatment
Temperature High
(Moderate)
X
Laser module treatment
temperature is high
Check fan and
ventilation. Allow system to
cool.
39
LBO Treatment
Temperature High
(Severe)
X
Laser module treatment
temperature is severely high
Check fan and
ventilation. Allow system to
cool.