<Spot Diameter> - press up/down buttons to adjust
B
<Spacing> - press up/down buttons to adjust
C
Primary Pattern - press to select pattern type
D
Secondary Pattern - press to select pattern subset (when applicable)
NOTE
•
Parameter controls only show on screens when applicable to the pattern selected.
•
Establish treatment parameters with Enhanced Octant when in “Standby” mode.
The Enhanced Octant pattern will increase the number of “spots” to accompany
changes in geometry (spot diameter, spacing, radii). Changes to geometry are
only possible when system is in “Standby” mode. Once in “Treat” mode, after the
first laser application is applied, only changes to Power, Exposure and Endpoint
are allowed.