Electromagnetic Compatibility
Applicable to PASCAL Synthesis Laser Systems with serial numbers starting with “80”
(eg 80xxxxxx).
This product conforms to the EMC standard (IEC 60601-1-2: 2014). The expected
electromagnetic environment for the whole life cycle is professional healthcare facility
environment.
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC information
provided in the ACCOMPANYING DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT.
c) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the EQUIPMENT or
SYSTEM should be observed to verify normal operation in the configuration in
which it will be used.
d) The use of the ACCESSORY, transducer or cable with EQUIPMENT and
SYSTEMS other than those specified may result in increased EMISSION or
decreased IMMUNITY of this EQUIPMENT and result in improper operation.
e) Do not use the devices generating electromagnetic waves within 30cm from all the
parts of the instrument and system. Those devices may have influence on this
product.
Item
Cable Shielded
Ferrite Core
Length (m)
ACCESSORIES
HEAD UP DISPLAY SYSTEM HUD-1 Module
-
-
-
CABLES
AC power cord (for SLIT LAMP)
Not Used
Not Used
1.8
AC power cord (for LASER CONSOLE)
Not Used
Not Used
3.6
USB Cable (Unterminated)
Used
Not Used
N/A
USB Cable (for 3D Mouse)
Used
Not Used
1.9
USB Cable (for Monitor)
Used
Not Used
1.9
VGA Cable (for Monitor)
Not Used
Not Used
1.9
SIP/SOP Cable (for Footswitch)
Not Used
Not Used
2.9
USB Cable (for HUD-1)
Not Used
Not Used
1.8
HDMI Cable (for HUD-1)
Not Used
Not Used
1.8