PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 52 of 142
88293-EN, Rev C
Array
Primary pattern
Secondary pattern
20ms Array
General Use
•
PRP, retina tears, retina detachments
Pattern
Spot Diameter (μm
)
Spacing
EPM %
20 ms Array
50 (2x2 or smaller)
0.00Ø to 3.00Ø
10 to 95
100
200
400
0.00Ø to 1.50Ø
The 20 ms Array pattern is selectable in a variety of shapes and sizes up to a maximum of
25 spots, including rectangle arrays (e.g., 2x3, 4x2, etc.), square arrays (e.g., 2x2, 3x3,
etc.), vertical and horizontal lines of up to five spots. Spot diameter and spacing settings
are also adjustable.
To select the Array pattern shape and size, drag your finger horizontally, vertically, or
diagonally across the pattern, or press one of the quick-select buttons at the bottom of the
Array pattern screen.
NOTE
•
Do not allow to swipe to single spot for all array patterns. If needed, please select
the single spot pattern.