Remove the pole caps (1) on the accessible battery (3).
•
Remove the pole screws located under the pole caps using an 11 mm socket spanner.
•
Remove the connecting cable (4) and the fuse cable (2).
•
Lift the battery from the battery carrier.
•
Pull on the strap of the second battery to move it forward.
•
Dismantle the cables on the second battery as well.
•
Lift the second battery from the battery carrier.
Refitting the batteries:
•
Refit the batteries in reverse order.
•
Ensure that the battery cage sockets/plugs have been correctly refitted. A polarity diagram is
located in the rear cover.