Hardware Manual: Part Number 10554 Rev C., iQue Operations: Daily Startup
Daily Startup
Overview
This procedure explains how to perform a proper daily startup. This entire procedure should take no longer than 20
minutes assuming the system was shut down properly and all the required fluids are prepared.
Referenced Procedures
•
•
•
Steps
1.
Ensure all bottles and troughs are filled according to the
procedure.
2.
Launch the ForeCyt software if it is not already running. When prompted to power on the detector, click
Yes
to
power it on. Alternately, click the
Power On Detector
button in the Controller Console window.
3.
Follow the Power On Detector wizard by disconnecting the tubing from the detector inlet and placing a tube of
filtered distilled water on the detector inlet.
4.
As the detector starts, the lasers and fluidics will go through a series of initialization steps which should take
about five minutes. It is possible to monitor the progress in the Controller Console window by clicking the
Show
Details
button. Once the detector is powered on, the system will show a status of Ready (green light).