Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Wetting Procedure
Running Water on Fast for 10 minutes
1.
Place a 5 mL tube containing distilled water on the detector inlet.
2.
Add another test tube to the experiment and set the Test Tube Time Limit to 10 minutes.
3.
Run the experiment.
Backflush
1.
Click the
Clean
button.
2.
In the Clean wizard, choose Clean Type =
Backflush
.
3.
Click
Next
and then
Start
to run the Backflush.
Unclog
1.
Click the
Clean
button.
2.
In the Clean wizard, choose Clean Type =
Unclog
.
3.
Click
Next
and then
Start
to run the Unclog.