iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Fluidics Maintenance
Fluidics Maintenance
Overview
The iQue detector has several parts that are designed to be regularly replaced in order to keep the system running in
optimal condition. These include various filters as well as the peristaltic pump tubes.
Frequency
At a minimum, fluidics maintenance should be performed every two months (bi-monthly). In a heavy usage situation,
this maintenance may need to be performed more frequently. See below for instructions on inspecting parts for wear
to determine if fluidics maintenance is overdue.
Required Materials
Part Number
Description
Works With
90075
iQue
®
/HTFC
®
Fluidics Maintenance Kit
Gen 1, Gen 2
90083
iQue
®
/HTFC
®
Fluidics Maintenance Kit 10
pack
Gen 1, Gen 2
Components in a Fluidics Maintenance Kit:
•
In-Line Sheath Filter
•
Peristaltic Pump Tubing
•
Sheath Bottle Filter
•
Decontamination/Cleaning Bottle Filters
Contents of the Fluidics Maintenance Kit
Summary of Contents for iQue SCREENER
Page 1: ...iQue SCREENER hardware manual October 2014...
Page 2: ......
Page 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Page 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Page 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Page 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 110: ......