![INTELLICYT iQue SCREENER Hardware Manual Download Page 42](http://html1.mh-extra.com/html/intellicyt/ique-screener/ique-screener_hardware-manual_2073911042.webp)
36
iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Operations: Daily QC
Creating the Daily QC Experiment
1.
In ForeCyt, click
File > New Experiment
.
2.
In the New Experiment dialog, select the option
Use Template
to create an experiment using a predefined
template.
3.
Click the
Browse
button and choose the Daily QC template that matches the bead lot being used.
NOTE
Contact IntelliCyt Tech Support to obtain the latest Daily QC Template for any specific
bead lot.
4.
Give the Daily QC experiment a name and click
OK
.
Running the Water Test
1.
Perform a Backflush by clicking the
Clean
button on the controller console, choosing the
Backflush
option, and
following the wizard.
2.
Place the Water Test sample on the detector inlet.
3.
Select the Water Test tube in the experiment.
4.
Click
Run
. The experiment will run for just over 2 minutes.